He then terminated the cycle by turning the power off, set the dip switches to invoke the "reset cycle to idle" process, turned the power back on, manually opened the door (while gas was in the chamber) and pressed the stop button (which has no effect after the reset cycle to idle procedure has been performed). He stated he tried to perform a "clear silent abort" as instructed on page 45 of the service manual but the sterilizer didn't respond. Reportedly, he was not wearing a monitor and there was no alarm in the room. Improper or Incorrect Procedure or Method (2017) Device Displays Incorrect Message (2591)ģm received information from a biomedical engineer that reported he was exposed to ethylene oxide (eo) when he was repairing a 3m steri-vac 5xl sterilizer that had stopped functioning and was displaying an e73 error. 3M HEALTH CARE 3M STERI-VAC 5XL STERILIZER/AERATOR 880.6860 STERILIZER, ETHYLENE-OXIDE GAS